Skin Analyzer Scores and Skin Age: How Far Can You Trust Them, and What Should You Compare Them With?

    465

Hello, I am Dr. Dong Won Kim, Medical Director of Zero Lab Clinic, one minute from Hongdae (Hongik Univ.) Station Exit 1.

It can be unsettling to get a skin analysis report back and see a low score. “My score came out low. Does that mean my skin is bad?” With numbers and even a skin age printed on it, it is only natural that it feels like a report card.

But these numbers are a different kind of value from blood test results. They are calculated by software from photographs, so what they mean depends on what they were calculated against and the conditions under which the photographs were taken. Today, using papers on the reproducibility and accuracy of facial imaging analysis devices, I will go through how far these numbers can be trusted and what they should be compared with.

How results are produced: photography → images under each light source → software analysis → absolute value, count, percentile and skin age

What does a skin analyzer photograph, and what does it calculate?

A facial imaging analysis device takes photographs under several kinds of light while the face rests against a fixed head support. Under ordinary white light, spots, wrinkles and skin texture are visible; under polarized or ultraviolet light, pigment and red areas that are hard to see with the naked eye show up separately. The software then finds features such as spots, wrinkles, pores and red areas in these photographs, counts these features and measures the area they cover.

Results usually come in three forms: the number of features, an absolute value combining area and intensity, and a percentile compared with data from the same age group. On top of this, some devices calculate and display a “skin age.”

Our clinic uses EVE V MUSE. It shoots with 5 cameras and 5 light sources and analyzes up to 34 items, including aging indicators, facial contours, pigmentation and pores.

First, I should be upfront about one point. We could not find any reproducibility study indexed in PubMed under the name EVE V MUSE. So the evidence in this article comes from papers on facial imaging analysis devices of the same type, mainly VISIA, the most extensively studied device. Because each device has different cameras and algorithms, the figures below cannot be transferred directly to our device. Please read them as evidence of how this type of device behaves as a whole.

If you retake it on the same machine, do you get the same score?

There is a study by a plastic surgeon in Germany, conducted independently of the device company. In it, 8 people were photographed 3 times in a row with their eyes closed and their faces against the fixed head support; 1 week later they came back, had their face position set again, and were photographed 3 more times.

The differences between the two sets of photos, a week apart, were under 2% for skin texture, UV spots, brown spots and porphyrins, 2–4% for pores and red areas, and around 6% for spots and wrinkles. When the researcher compared the repeated photographs by eye, no difference could be made out either, except in the red areas of 1 person[1].

In another study by the same researcher, in which 19 people were rephotographed 4 weeks apart, the standard deviation of the absolute wrinkle value was likewise 3.36% the first time and 3.4% the second[2].

With the same machine, the same head support and the same posture, the numbers are fairly stable. However, how much they fluctuate differs by item. If a wrinkle or spot score moves by a few percent, it is hard to conclude that your skin changed in the meantime. And both were small studies, of 8 and 19 people.

ItemDifference on retake, same machineEvidence
Skin texture, UV spots, brown spots, porphyrinsUnder 2%[1]
Pores, red areas2–4%[1]
Spots, wrinklesAbout 6%[1]
Wrinkles, absolute valueStandard deviation about 3%[2]
Wrinkles, percentileStandard deviation about 9%[2]

If my “X% compared with my age group” score is low, does that mean my skin is bad?

The percentile, the number that catches the eye first on the report, shows roughly where you would rank among 100 people your age in the device’s reference data. But in the 4-week retake study above, the standard deviation of the percentile was 8.2% the first time and 10.7% the second, nearly 3 times larger than that of the absolute value (about 3%). The researcher concluded that the more precise absolute value should be looked at first[2].

The percentile is a comparison with age-group data held by the device company, so the same skin can rank differently depending on how that data is composed. This time we could not find any material that separately verified percentile accuracy for Asian skin.

That is why we do not judge skin to be bad simply because the percentile came out low. The percentile is a reference value; to see how your own skin has changed, you need to look at how the absolute value for the same item has moved since last time.

Comparison placing side by side the standard deviation on retaking the same person: absolute value about 3%, percentile about 9%

Do scores change with lighting, angle, makeup or season?

They do. Analyzer scores come from photographs, so when the photograph changes, the number changes too.

Angle and distance. In a study that photographed standard skin-color cards with a smartphone while varying distance and angle, angle changed both brightness and hue, and distance changed brightness. Turning on the flash generally did not reduce this variation[3]. The experiment photographed color cards rather than faces, but it is evidence that if the shooting angle is not consistent, skin photographs taken over time can look different.

Makeup. When a face wearing foundation is photographed with UV fluorescence imaging, how much foundation remains shows up directly in the image. The correlation coefficient between the amount wiped off and the image value was -0.86[4]. This study was carried out at a cosmetics company’s laboratory. This means that if you are photographed with makeup or sunscreen still on, the pigment and UV items may reflect the product you applied rather than your skin.

Season. In a study that measured 100 Korean men aged 20–59 in summer and in winter, using a Korean facial analysis device, significantly more pores were detected in summer than in winter. Sebum was also higher in summer, while wrinkles and skin brightness showed no seasonal difference[5].

Time of day. In a study that repeatedly measured the skin of healthy Chinese women over the course of a day, sebum secretion, water loss, hydration and skin temperature rose and fell on a daily cycle[6]. The measurements were made with skin-measuring instruments rather than an imaging device, but it is hard to regard morning and evening results as taken under the same conditions.

If you directly compare last summer’s result with this winter’s, you may be looking at the difference between seasons rather than a difference in your skin. The pore score is one such item.

Checklist of conditions to match before comparing results: same machine, same posture, bare face, similar time of day, similar season

Can I compare my score with one from another clinic’s machine?

On this question I think I have to answer somewhat firmly. If the devices differ, the scores are hard to compare directly.

In a study of 28 people in which the same 2×2 cm area was photographed with two types of device, the brown spot and red area items correlated between the two devices. But wrinkles and pores showed no linear correlation between the two devices. The UV spot score from one device also did not correlate with the melanin value from the other[7].

In a study that analyzed 50 people by two methods, a tablet app and a photo booth, the two devices had an agreement rate of 67.7%. By item, that was skin texture 72.0%, pores 68.2%, wrinkles 67% and redness 64%[8].

In a study that measured facial pigmentation in 80 people with two devices, too, the correlation coefficients between the two devices’ readings were 0.331 and 0.395, which is on the low side[9].

Even under the same name, “wrinkles,” each device differs in what it counts as a wrinkle and what lighting it shoots under. Even for items with the same name, you should treat numbers from different devices as values in different units.

Bars for the agreement rate between two devices in percent: skin texture 72.0, pores 68.2, wrinkles 67, redness 64

How well do machine scores match what a doctor sees by eye?

In the 80-person study above, 3 dermatologists graded pigmentation by eye, and their grades were compared with the two devices’ readings. The correlation coefficients were 0.509 and 0.653, which is moderate[9].

There is also a 2026 study that built an AI skin analysis model from facial photographs of 1,099 Koreans. Compared with the grades assigned by 5 dermatologists, the model got the grade exactly right an average of 51.4% of the time and came within 1 grade 93.3% of the time. In photographs taken with a tablet or personal smartphone instead of a high-end camera, accuracy for pigment-related items dropped noticeably[10]. One of the authors is affiliated with a company.

Machines and doctors generally point in the same direction, but differences of around 1 grade are common. So we do not decide on a procedure from an analyzer score alone.

It is most meaningful when the same person is compared over time

This is the important part. The analyzer’s numbers are most useful when the same person is compared on the same device over time.

There is a study that analyzed photographs of 227 patients with flushing-type rosacea, taken on the same device before and after 3 sessions of intense pulsed light (IPL) treatment. In it, 3 dermatologists separately rated the degree of redness from the photographs alone. The reduction in red area correlated with the improvement rated by the dermatologists (correlation coefficient 0.476), and the area under the curve (AUC) for distinguishing patients whose redness improved by 2 grades or more was 0.851[11].

However, this was a retrospective single-center study, and the authors themselves wrote that the cutoff of 18.76% they proposed is an exploratory value that needs to be confirmed at other institutions.

A change measured against your last result tells you more than any one score taken on its own. That holds on the condition that the photographs were taken on the same machine under the same conditions.

How accurate is “skin age”?

In the 19-person study above, the correlation coefficient between the skin age calculated by the device and actual age was 0.896 for the right side of the face and 0.827 for the left, which is high. On average, skin age came out 1.37 years older than actual age. But looking at each person individually, the estimated skin age was within 3 years of actual age for half of the participants, and within 4.5–5.5 years for 3 out of 4[2].

This means the group average fits well, but for any one person an error of about 3–5 years in skin age is common. Results for the right and left sides of the same face also differed. And this study was of 19 people in Germany. It cannot tell us how accurate skin age is for Asian skin.

Skin age is a value calculated from visible features such as wrinkles and pigmentation. It is not a test that measures the health of the skin or the speed of aging.

Number cards for skin age and actual age: correlation 0.896, average difference 1.37 years, half within 3 years, 3 out of 4 within 4.5–5.5 years

What we check at consultation

  • Whether you have a previous result — if so, we explain based on how the absolute value for the same item has moved
  • Today’s photo conditions — whether your face is bare after cleansing, or whether sunscreen or makeup remains
  • The seasonal gap from your last photo — when comparing summer and winter results, we look at and explain the pore and sebum items separately
  • Whether you brought a report from another clinic — if the device is different, we do not carry the numbers over for comparison, and refer only to what can be seen in the photographs
  • What you want to check with the analyzer — the photographs and items we look at differ depending on whether it is pigmentation, redness or wrinkles

We sometimes take analyzer photographs at the first consultation and sometimes do not. When comparing before and after a procedure, it varies by procedure, but we usually retake them after about 2 weeks, and we can print the report for you or send it by email.

Frequently asked questions

Can I compare my score with one taken at another clinic?

If the devices differ, direct comparison is difficult. In a study that photographed the same area with two devices, wrinkles and pores showed no linear correlation between the two devices, and in another study the agreement rate between two devices was 67.7%.

If my skin age comes out higher than my actual age, does that mean I am aging fast?

That is hard to conclude. In one study, skin age correlated highly with actual age, but for individuals, the estimate was within 3 years of actual age for half of the participants and within 4.5–5.5 years for 3 out of 4. It is a value calculated from visible features, so it is not a test that measures the speed of aging.

I retook it a month later and the score changed a little. Has my skin changed?

Even when retaken on the same machine a week later, differences of under 2% to about 6% appear depending on the item. There is also a study in which pore scores came out higher in summer. A small change may be due to the photo conditions or the season.

What are analyzer scores most useful for?

Comparing the same person on the same device over time. In a study that photographed patients before and after IPL treatment on the same device, the reduction in red area correlated with the improvement rated by dermatologists.

In closing

Many people ask whether a low analyzer score means their skin is bad. It is hard to judge skin as bad from one score alone.

Retaken on the same machine, the numbers are fairly stable, but the percentile compared with your age group varies between shots by nearly 3 times as much as the absolute value. The numbers change with makeup, season and shooting angle, and on different devices even the same item can represent a different value. Skin age fits well on average, but individual errors of 3–5 years are common.

We have not yet found a study under our device’s name, and there is the limitation that I have spoken from evidence on devices of the same type, but the message I most want to get across is that an analyzer score means the most not as a report card for comparing yourself with others, but as a record for comparing yourself with your last result. So when you get your report, I recommend looking at the photographs together first, before the numbers.


References

[1] Henseler H. Investigation of the precision of the Visia® complexion analysis camera system in the assessment of skin surface features. GMS Interdiscip Plast Reconstr Surg DGPW. 2022;11:Doc08. DOI: https://doi.org/10.3205/iprs000169

[2] Henseler H. Assessment of the reproducibility and accuracy of the Visia® Complexion Analysis Camera System for objective skin analysis of facial wrinkles and skin age. GMS Interdiscip Plast Reconstr Surg DGPW. 2023;12:Doc07. DOI: https://doi.org/10.3205/iprs000177

[3] Cronin A, Tkaczyk ER, Hussain I, et al. Effect of camera distance and angle on color of diverse skin tone-based standards in smartphone photos. J Biophotonics. 2023;16(6):e202200381. DOI: https://doi.org/10.1002/jbio.202200381

[4] Messaraa C, Walsh M, Hurley S, et al. A novel UV-fluorescence approach to assess the long wear efficacy of foundations. Skin Res Technol. 2021;27(5):758-765. DOI: https://doi.org/10.1111/srt.13011

[5] Song EJ, Lee JA, Park JJ, et al. A study on seasonal variation of skin parameters in Korean males. Int J Cosmet Sci. 2015;37(1):92-97. DOI: https://doi.org/10.1111/ics.12174

[6] Lv L, Yan X, Zhou M, et al. Circadian rhythms of skin surface lipids and physiological parameters in healthy Chinese women reveals circadian changes in skin barrier function. Biology (Basel). 2024;13(12):1031. DOI: https://doi.org/10.3390/biology13121031

[7] Linming F, Wei H, Anqi L, et al. Comparison of two skin imaging analysis instruments: The VISIA® from Canfield vs the ANTERA 3D® CS from Miravex. Skin Res Technol. 2018;24(1):3-8. DOI: https://doi.org/10.1111/srt.12381

[8] Cook MK, Kaszycki MA, Richardson I, et al. Comparison of two devices for facial skin analysis. J Cosmet Dermatol. 2022;21(12):7001-7006. DOI: https://doi.org/10.1111/jocd.15434

[9] Zuo Y, Li A, He H, et al. Assessment of features in facial hyperpigmentation: Comparison study between VISIA and CSKIN. Skin Res Technol. 2022;28(6):846-850. DOI: https://doi.org/10.1111/srt.13216

[10] Lee E, Lee J, Kim N, et al. Artificial intelligence based skin analysis models for predicting visual grades and device measured physiological values from facial images. Skin Res Technol. 2026;32(9):e70375. DOI: https://doi.org/10.1111/srt.70375

[11] Fan B, Meng X, Yu R. Evaluation of VISIA-derived erythema metrics against blinded clinician assessment after intense pulsed light therapy in erythematotelangiectatic rosacea: a real-world cohort study. Clin Cosmet Investig Dermatol. 2026;19:649214. DOI: https://doi.org/10.2147/CCID.S649214

The 11 items above are papers indexed in PubMed. This article is general medical information; decisions about an individual’s condition are made through consultation.